Installation and Operation Manual
X-TMF-GF40-GF80-Series-MFC-eng
Part Number: 541B161AAG
June, 2013
Section 2 Installation
2-11 Purge the Gas Supply Line Before GF40/GF80/GF81 Series Installation
Before operating the GF40/GF80/GF81 Series, the gas supply line must be
completely purged with nitrogen or argon to ensure the line is free from toxic
or flammable gases, contaminants, moisture, and oxygen. The purge gas
must be free of moisture and oxygen to less than 100 ppb. Purge the gas
lines as follows or in accordance with prescribed company and safety
procedures.
1. Shut off the process gas supply valve(s) upstream of the GF40/GF80/
GF81 Series. If such a valve is not available, shut the valve on the gas
panel. Tag the valve at this point to prevent accidental re-exposure of the
process gas to the gas line.
2. Cycle purge the gas line with dry nitrogen or argon to fully flush out the
process gas. Cycle purging consists of evacuating to a low pressure
adequate to induce out-gassing and then purging to remove adhered
moisture and oxygen. If a toxic or reactive gas is present and a clogged
GF40/GF80/GF81 Series is suspected, then proceed with caution.
Pump down and purge the GF40/GF80/GF81 Series from both down-
stream and upstream lines. If check valves are present in the gas line,
both pumping down and purging are required. Pumping down without
purging is inadequate. If a good vacuum source is not available, the
GF40/GF80/GF81 Series can be de-contaminated by purge only.
3. Repeat the purge cycle several times within 2-4 hours to complete the
cleaning. For toxic and corrosive gasses, it is recommended to use 100-
120 cycles.